|
|
|
|
|
|
|
A two-photon microscope is similar to a confocal microscope. Both provide optical sectioning and are very well suited for 3D imaging. In both systems a laser beam is focused to a small spot and scanned line-by-line over the sample.
In a confocal microscope, all dye molecules along the beam-path are excited (and eventually bleached). Out of focus light is rejected using a pinhole.
In a two-photon microscope, only dye molecules within the focal spot are excited, thus there is no need for a pinhole. This fact and the use of near-infrared-light (NIR) allows imaging of thicker biological samples.
Two-photon microscopes use near-infrared (NIR) light to excite dyes normally excited with shorter wavelengths in one-photon microscopy. Most tissue is transparent to NIR light. NIR light is reported to be less harmful to live cells and developing animals.
The two-photon effect can be described in terms of non-linear optics (NLO). The terms NLO, two-photon microscope or multi-photon microscope (MPM) are often used interchangeably, even if their meanings are not identical.
At the Stowers Institute we have three two-photon microscopes.
|
|
|
|
Stowers Links:
Image Center: Instruments
Adv. Instr. & Physics: Confocal Laser Scanning Microscope (LSM)
Foundations of Microscopy Seminar: Non-Linear Optics
Imaging in 3D: Confocal vs. NLO Microscopy: View Presentation/Download Presentation (zip)
Web Links:
Molecular Expressions: Multiphoton Fluorescence Microscopy
User group: Multi-Photon Laser Scanning Microscopy
W.W.Webb Group at Cornell University
Literature:
Molecular Expressions: Multiphoton Fluorescence Microscopy - Selected Literature
Denk, W., J. H. Strickler, et al. (1990). "Two-Photon Laser Scanning Fluorescence Microscopy." Science 248(4951): 73-76.Download PDF
M. Dickinson: Multiphoton Laser Scanning Microscopy: Introduction into Theory and Operation
Koenig, K. (2000). "Multiphoton microscopy in life sciences." Journal of Microscopy, 200(2): 83-104Download PDF
Mertz, J. (2004). "Nonlinear microscopy: new techniques and applications." Curr Opin Neurobiol 14: 610-6.Download PDF
|
|
|
|
|
|
Download Presentation |
| Go to link page |
|
|
|
|
|
|
|
|
|
|
|
|
In one as well as two-(or multi-) photon fluorescence, a dye molecule emits light when an electron relaxes from an excited energy state to it's ground level. The energy difference between the two levels (electronic and vibrational states) determine the emission spectrum. It is identical for one- and multi-photon excitation.
The electron moves from the ground level to the excited level by absorbing one or multiple photons. Typically this state is a vibrational state and the electron dissipates energy before emitting a photon and relaxing to the ground level. The sum of the energies of all absorbed photons has to be larger than the energy gap and is larger (or equal) to the energy of the emitted photon.
In one photon fluorescence, this translates into the absorbance of one photon with a shorter wavelength than the emitted photon.
In two-photon fluorescence two photons with about double the wavelength (half the energy) are absorbed.
|
|
|
|
Stowers Links:
Adv. Instr. & Physics: Fluorescence
|
|
|
|
|
|
Download Presentation |
| Go to link page |
|
|
|
|
|
|
|
|
|
|
|
|
For two-photon excitation, two photons have to be absorbed by the dye molecule at the same time
(within about 15 fs). This requires an extreme high photon flux, only present in the focus of a high numerical aperture lens using a strong, pulsed near infrared laser. This is shown in the image. In the upper part of the curvet, all molecules within the beam-path are excited, while in the lower part only molecules within the focal spot emit light.
The optical sectioning capability of a two-photon microscope is based on this fact - only dye molecules within the in-focus plane are excited. Thus, contrary to a confocal microscope, a two-photon microscope does not need a pinhole.
|
|
|
|
|
|
|
|
|
|
Download Presentation |
| Go to link page |
|
|
|
|
|
|
|
|
|
|
|
|
Two-photon excitation follows different selection rules than one-photon excitation. Thus one can not just shift the one-photon excitation spectrum to twice the wavelength. Typically two-photon spectra are shifted to shorter wavelengths than expected by doubling the wavelength of one-photon spectra. Two-photon spectra are in most cases much broader than one-photon spectra.
Good results have been reported for the following excitation wavelengths:
Blue/Cyan Dyes
Dye |
Excitation |
| Alexa 350 |
780-800 nm |
| Hoechst |
780-800 nm
900-1100 nm
|
| DAPI |
780-800 nm
900-1100 nm
|
| CFP |
800-900 nm |
Green Dyes
Dye |
Excitation |
| Oregon Green |
800-860 nm |
| Alexa 488 |
800-830 nm
|
| eGFP |
920-990 nm
|
| BODIPY |
900-950 nm |
| FITC |
750-800 nm |
| DiO |
780-830 nm |
Yellow/Orange Dyes
Dye |
Excitation |
| YFP |
890-950 nm |
| DiA |
800-860 nm
|
Red Dyes
Dye |
Excitation |
| DiI |
830-920 nm |
| Rhodamine B |
800-860 nm
|
| Alexa 568 |
780-840 nm
|
Unfortunately not many two-photon excitation spectra are published. Reported excitation wavelengths are often unreliable. We typically use excitation fingerprinting to determine the best imaging conditions.
|
|
|
|
Web Links:
Carl Zeiss CellScience: Multi-photon Spectra
Cornell University (Ward W. Webb): 2P cross sections
Literature:
Fisher, J. A. N., B. M. Salzberg, et al. (2005). "Near infrared two-photon excitation cross-sections of voltage-sensitive dyes." Journal of Neuroscience Methods 148(1): 94-102.Download PDF
Xu, C., W. Zipfel, et al. (1996). "Multiphoton Fluorescence Excitation: New Spectral Windows for Biological Nonlinear Microscopy." PNAS 93(20): 10763-10768.Download PDF
|
|
|
|
|
|
Download Presentation |
| Go to link page |
|
|
|
|
|
|
|
|
|
|
|
|
Most animal cells and tissues are considered to be nearly transparent in the spectral range from about 700 nm to 1100 nm owing to the lack of efficient one-photon absorbers.
This enables deeper penetration of the NIR
excitation laser-beam into tissue. Damage to the sample is reduced.
The graphic shows absorption spectra of major intracellular absorbers. The molecular exctinction coefficients of oxygenated haemoglobin and melanin and the absorption coefficient of water are shown.
|
|
|
|
Literature:
Koenig, K. (2000). "Multiphoton microscopy in life sciences." Journal of Microscopy, 200(2): 83-104Download PDF
|
|
|
|
|
|
Download Presentation |
| Go to link page |
|
|
|
|
|
|
|
|
|
|
|
|
Most components of a confocal laser scanning microscope and a two-photon system are identical. All two-photon microscopes at the Stowers Institute can be used in addition as confocal microscopes with excitation wavelengths in the visible range.
The pulsed NIR laser light is coupled into the confocal detection module. It is directed toward the sample using a short-pass filter (e.g. KP 700). Two mirrors scan the laser-beam over the sample. The detection head is attached to a standard microscope. The beam is focused into the sample with a high numerical aperture lens optimized for near-infrared transmission and minimum dispersion.
Fluorescence is collected using the identical lens. The fluorescence light hits again the scanning mirrors. Depending on the position of the scanned excitation laser beam, the fluorescence is generated at different locations inside the sample. After the fluorescence light passes the scanning mirrors, it is stable. This beam path is called the de-scanned detection path. Because the fluorescence in a two-photon setup has a shorter wavelength than the excitation laser-beam, it can pass the short-pass filter. The light passes through the pinhole. In most two-photon applications the pinhole is opened to it's maximum size. The light is detected using one or multiple photo-detectors.
The de-scanned detection path is not the most efficient one to detect light scattered out of focus. It is more efficient not to route the fluorescence light back into the detection head and not to let it pass the scanning mirrors. A short-pass filter close to the sample (typically in the filter turret of the light microscope) reflects the fluorescence light directly onto a detector. Because now the light is moving in synchrony with the excitation laser beam, the detector has to be large enough and no pinhole can be used.
In some cases forward fluorescence or transmitted light is collected using the condenser of the microscope.
Two-Photon Microscopes at Stowers Institute
System |
|
|
Ron Yu Lab |
Location |
451 G Bldg 2 |
461 Bldg 2 |
1xx Bldg 1 |
Excitation |
NIR |
720-950nm |
690-1040nm |
690-1040nm |
|
458nm |
x |
x |
x |
|
488nm |
x |
x |
x |
|
514nm |
x |
x |
x |
|
543nm |
x |
- |
- |
|
561nm |
- |
x |
x |
|
633nm |
x |
x |
x |
| Microscope |
Axiovert 200 (inverted) |
Axiovert 200 (inverted) |
Axioskop 2FS (upright, physiology) |
Software |
AIM 3.2 |
AIM 4.0 |
AIM 4.0 |
NDD |
x |
x |
- |
| Additional Equipment |
Environmental chamber |
FCS
FLIM |
Patch clamp |
|
|
|
|
Stower's Links:
Adv. Instr. & Physics: Confocal Laser Scanning Microscope (LSM)
Image Center: Instruments
Ron Yu Lab
Web Links:
Carl Zeiss Laser Scanning Microscopes
Literature:
Manual LSM 510 META 3.5
|
|
|
|
|
|
Download Presentation |
| Go to link page |
|
|
|
|
|
|
|
|
|
|
|
|
To achieve the photon density necessary for two-photon excitation pulsed near-infrared lasers are used. Typically they are diode pumped Ti:Sapphire lasers. At the Stowers Institute we are using Coherent Chameleon lasers. The wavelength can be tuned and the laser is controlled by the microscope software.
Coherent Chameleon Pulsed NIR lasers
| Parameter[1] |
Specifications |
Chameleon |
Chameleon ULTA |
System[2] |
LSM 510-NLO |
LSM 510-DEV
Ron Yu lab |
Average Power |
> 1W |
> 2.0W |
Pulse Width[3] |
< 140 fs at peak
< 200 fs across the tuning range |
Repetition Rate |
90 MHz |
80 MHz |
Tuning Range[4] |
720 to 950 nm |
690 to 1040 nm |
Power Stability[5] |
<±0.5% |
Noise[6] |
<0.15% rms |
M2 (Beam Quality) |
<1.1 (TEM00) |
[1]All specifications apply at peak of tuning curve unless otherwise stated.
[2]Systems at the Stowers Institute
[3]Full width at half maximum
[4]Wavelength accuracy is 2nm
[5]Power drift specified in any two-hour period with less than ±1°C temperature change after one-hour warm-up.
[6]Measured from 10Hz to 20MHz
|
|
|
|
|
Web Links:
Coherent Chameleon Ultrafast Ti:Sapphire Lasers
Molecular Expressions:
Ti:Sapphire Mode-Locked Lasers
Literature:
Coherent Chameleon Ultra : Manual. Chapter 7 is a good introduction into the physics of pulsed lasers.
Chameleon Data sheet Download PDF
Chameleon Ultra Data sheet Download PDF
|
|
|
|
|
|
Download Presentation |
| Go to link page |
|
|
|
|
|